GRINjector is a device for injecting gradient-refractive index (GRIN) lens probes into animals. The advantages are:

  1. The injection needle displaces brain tissue, improving targetting and minimizing tissue compression.
  2. The GRIN is held in place during retraction of the placement needle from the GRIN and the GRINJector device.
  3. The guide needle can be easily swapped for other guide needles depending on the diameter and length of the GRIN lens probe.

This repository contains CAD files for printing GRINjector version 3.0 components and assembling them into a usable GRINjector.

For those conducting calcium imaging experiments and analysis, we have a software package called CIAtah: https://github.com/bahanonu/ciatah.

Questions? Contact: Biafra Ahanonu (github [at] bahanonu [dot] com).

Components

The below figure illustrates several of the main components:

image

Below is a list of components needed to assemble a GRINjector:

Printing

The main parts are provided as STEP and STL files to allow printing on nearly all commonly used 3D printers (Straysys, uPrint, ). We recommend printing with a high-resolution printer to ensure optimal fit and one with easy to remove support material—we have used VeroBlackPlus on a Stratasys Objet30 and found that to work quite well.

Assembly

Below are step-by-step by step instructions for assembling the GRINjector after building the necessary parts.

  1. Connect the SM-13 and 9066-COM to form a functional setup. Remove the locking plate from 9066-COM as below (see https://www.newport.com/medias/sys_master/images/images/he5/h5b/9366526427166/906806B-9066-9067-User-Manual.pdf).
<p align="center"><img src="https://user-images.githubusercontent.com/5241605/97954746-3a291d80-1d59-11eb-84a5-520f62bd6339.png" width="90%"></p>
  1. Using 8-32 cap screws, connect the Stereotaxic holder attachment to the back of the 9066-COM, connecting to the top-most holes near the micrometer.
<p align="center"><img src="https://user-images.githubusercontent.com/5241605/98111550-9bceb200-1e55-11eb-8d65-08629b0a189f.png" width="30%"></p>
  1. Next, pull down the bearing stage so that the two 4-40 holes in the back are exposed, place 4-40 x 5/16'' cap screws in each hole.
<p align="center"><img src="https://user-images.githubusercontent.com/5241605/98112684-64f99b80-1e57-11eb-8d2b-5b6605a19908.png" width="30%"></p>
  1. Tap the grinjector_mainneedle_connector_bottom holes with a 4-40 tap then place it onto 9066-COM as below and screw in the 4-40 x 5/16'' cap screws (red below).
<p align="center"><img src="https://user-images.githubusercontent.com/5241605/98113833-0f25f300-1e59-11eb-8dc2-21cbd99d3b28.png" width="30%"></p>
  1. Align the lock screws that comes with 9066-COM to each of the four holes on each side of the grinjector_bottom_connector.
<p align="center"><img src="https://user-images.githubusercontent.com/5241605/98114951-e0108100-1e5a-11eb-87b9-db21052ec571.png" width="70%"></p>
  1. Next, place a blunt or syringe tip Holder needle that matches the size of your GRIN lens probe into the groove of the grinjector_mainneedle_connector_bottom piece then place the grinjector_mainneedle_connector_top piece on top. Screw the grinjector_mainneedle_connector_top in place with 4-40 x 5/16'' cap screws then secure the Holder needle with a 4-40 x 1/4'' cap screw.
<p align="center"><img src="https://user-images.githubusercontent.com/5241605/98116044-92951380-1e5c-11eb-823e-143962dddfa5.png" width="40%"></p>
  1. Next, obtain a long, Stability needle (e.g. ~80-mm long, 0.5-mm diameter) and thread through the small opening at the top of grinjector_bottom_connector. The Stability needle will then go through the Holder needle. Lock the needle in place with a 4-40 x 3/16'' cap screw.
<p align="center"><img src="https://user-images.githubusercontent.com/5241605/98115791-3c27d500-1e5c-11eb-88a2-a93a26b4b4f6.png" width="30%"></p>
  1. Lastly, verify that the chosen Holder needle fits the GRIN lens probe tightly and that you can push the GRIN lens probe out of the Holder needle smoothly by twisting the SM-13 micrometer.
<p align="center"><img src="https://user-images.githubusercontent.com/5241605/98116361-0505f380-1e5d-11eb-9d98-f1d25e4e50c8.png" width="60%"></p>

You are now done assembled the GRINjector! Happy imaging.

<p align="center"><img src="https://user-images.githubusercontent.com/5241605/98117424-93c74000-1e5e-11eb-941a-dd73a9046b34.png" width="40%"></p>

Animal experiments

To use the GRINjector in an animal:

<p align="center"><img src="https://user-images.githubusercontent.com/5241605/98118465-1a305180-1e60-11eb-9ea6-34a963d214c3.png" width="100%"></p>

Notes

Holder needle:

Stability needle:

Acknowledgements

I would like to thank Prof. Benjamin Grewe for helping inspire and provide feedback on the initial GRINjector design.

References

We have a book chapter pending describing this device. In the meantime, if you used this device in your research please cite our publication Corder*, Ahanonu*, et al. Science 2019 :

@article{corderahanonu2019amygdalar,
  title={An amygdalar neural ensemble that encodes the unpleasantness of pain},
  author={Corder, Gregory and Ahanonu, Biafra and Grewe, Benjamin F and Wang, Dong and Schnitzer, Mark J and Scherrer, Gr{\'e}gory},
  journal={Science},
  volume={363},
  number={6424},
  pages={276--281},
  year={2019},
  publisher={American Association for the Advancement of Science}
}